Accurate and simple determination of oxcarbazepine in human plasma and urine samples using switchable-hydrophilicity solvent in GC-MS
[ X ]
Tarih
2020
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Wiley
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
This work presents a sensitive and rapid analytical method for the determination of oxcarbazepine in human plasma and urine samples. A vortex-assisted switchable hydrophilicity solvent-based liquid phase microextraction (VA-SHS-LPME) was used to preconcentrate oxcarbazepine from the samples before the determination by gas chromatography mass spectrometry. The switchable hydrophilicity solvent was synthesized by protonatingN,N-dimethylbenzylamine with carbon dioxide to make it totally miscible with an equivalent volume of water. Parameters of the VA-SHS-LPME method including volume of switchable hydrophilicity solvent, concentration/volume of sodium hydroxide and vortex period were systematically optimized. Under the optimum conditions, good linearity ranging from 27.03 to 353.47 mu g/kg was obtained for the analyte. Limit of detection and quantitation values were found to be 6.2 and 21 mu g/kg (mass base), respectively. The relative standard deviation was calculated as 6.9% for six replicate measurements of the lowest concentration of the calibration plot. Satisfactory recovery results were calculated in the range of 97-100% for human plasma and urine samples spiked at five different concentrations.
Açıklama
Anahtar Kelimeler
gas chromatography-mass spectrometry, human plasma, oxcarbazepine, switchable hydrophilicity solvent, urine
Kaynak
Biomedical Chromatography
WoS Q Değeri
Q4
Scopus Q Değeri
Q3
Cilt
34
Sayı
10